Add to Cart
Medical Titanium Rod:
Specifications: | Rolling ¢8.0---40mm× L; Forging ¢40---150mm×L |
Metallurgical structure: | Pure titanium grain size is not less than grade 5, TC4 titanium alloy meets A1-A9. |
Surface: | Black leather surface, polished surface, polished surface (H11, H9, H8) |
Performance of medical titanium rods | reference implementation standards: GB/T13810-2007, ASTM F67/F136 |
Room temperature properties of medical titanium rods GB/T13810-2007:
Grade
|
Status
|
Diameter or side length
|
Tensile strengthσb
|
Yield Strengthσ0.2
|
Elongation after fracture
|
Rate of reduction in area
|
mm
|
MPa
|
MPa
|
A%
|
Z%
|
||
TA1ELI
|
M
|
>7~90
|
≥200
|
≥140
|
≥30
|
≥30
|
TA1
|
M
|
≥240
|
≥170
|
≥24
|
≥30
|
|
TA2
|
M
|
≥400
|
≥275
|
≥20
|
≥30
|
|
TA3
|
M
|
≥500
|
≥380
|
≥18
|
≥30
|
|
TA4
|
M
|
≥580
|
≥485
|
≥15
|
≥25
|
|
TC4
|
M
|
>7~50
|
≥930
|
≥860
|
≥10
|
≥25
|
M
|
>50~90
|
≥895
|
≥830
|
≥10
|
≥25
|
|
TC4ELI
|
M
|
>7~45
|
≥860
|
≥795
|
≥10
|
≥25
|
M
|
>45~65
|
≥825
|
≥760
|
≥8
|
≥20
|
|
M
|
>65~90
|
≥825
|
≥760
|
≥8
|
≥15
|
|
TC20
|
M
|
>7~100
|
≥900
|
≥800
|
≥10
|
≥25
|
Note: For titanium bars with a diameter greater than 75 mm, take the bar-shaped specimen.
|
The mechanical properties of medical titanium rods comply with GB/T13810-2007 and ASTM F67/F136.
Using titanium rods to make artificial bones, cells can regenerate on them and bones can grow. Titanium has good affinity with human bones, epithelium, and artificial tissue. In order to improve the surface quality of implants connected to human bones, applying a bioceramic layer to the roughened surface is also a more effective method.